Diagnostics and troubleshooting Liquistation CSF48
212 Endress+Hauser
Vacuum Read only The vacuum is an indicator of the suction height.
-> 100 mbar is equivalent to a suction height of approx.
1 m.
Medium detected Read only Yes: the medium was detected
No: no medium was detected
Pump suction Action
Pump suction, to stop
press ESC
Read only
Current pump run time Read only
Power supply Read only The current supply voltage is displayed.
With AC power supply: 24 V ±0.5 V
With DC power supply: 22 to 28 V
Motor current Read only The current consumption of the pump is displayed.
Vacuum Read only The vacuum is an indicator of the suction height.
-> 100 mbar is equivalent to a suction height of approx.
1 m.
Medium detected Read only Yes: the medium was detected
No: no medium was detected
Vacuum pump
(only for version with vacuum
pump)
Action
Bottle configuration Read only
Bottle volume Read only
Distributor position Options
•Front
•Bottle 1
...
•Back
Select which bottle should be filled with the sample.
Sample volume Factory setting
200 ml
The sample volume is preset at the factory.
Start sampling Action Perform sampling manually.
Progress Read only The progress of the sampling operation is displayed.
Power supply Read only The current supply voltage is displayed.
With AC power supply: 24 V ±0.5 V
With DC power supply: 22 to 28 V
Motor current Read only The current consumption of the pump is displayed.
Diagnostics/Systemtest/Reset
Function Options Info