PROline Promass 80 5Operation
Endress+Hauser 47
17
0 not assigned
1 not assigned
2 not assigned
3 not assigned
4 not assigned
5 not assigned
6 not assigned
7 671 Simulation status input active
18
0 not assigned
1 not assigned
2 not assigned
3 691 Simulation of response to error (outputs) active
4 692 Simulation of measured variable active
5 not assigned
6 not assigned
7 not assigned
19
0 700 The process fluid density is outside the upper or lower limit values
set in the “EPD” function
1 701 The maximum current value for the measuring tube exciter coils
has been reached, since certain process fluid characteristics are
extreme.
2 702 Frequency control is not stable, due to inhomogeneous fluid.
3 703
Overdriving of the internal analog to digital converter.
A continuation of the measurement is still possible!
4 704
5 705 The electronics' measuring range will be exceeded.
The mass flow is too high.
6 not assigned
7 not assigned
20
0 not assigned
1 not assigned
2 not assigned
3 not assigned
4 not assigned
5 731 The zero point adjustment is not possible or has been cancelled.
6 not assigned
7 not assigned
Byte Bit Error No. Short error description (Page 62 ff. )