Description of Device Functions Proline Prosonic Flow 93T Portable
104 Endress+Hauser
Function group ADJUSTMENT
Function group PIPE DATA
Functional description
BASIC FUNCTION PROCESS PARAMETER ADJUSTMENT
ZERO POINT ADJUSTMENT
(6480)
Use this function to start zero point adjustment.
!
Note!
For information on how to perform a zero point adjustment, see ä 44.
Options
CANCEL
START
Factory setting
CANCEL
Functional description
BASIC FUNCTION PROCESS PARAMETER PIPE DATA
PIPE STANDARD
(6520)
Use this function to select a pipe standard.
Options
OTHERS
DIN: PN10, PN16, 28610, 28614, 28615, 28619
ANSI: SS SCH40S, SS SCH80S, SS SCH5S, SS SCH10S
CS SCH20, CS SCH40, CS SCH80, CS SCH120,
AWWA: CLASS 50, CLASS 53, CLASS 55
!
Note!
The selection specifies the values for the following functions:
• PIPE MATERIAL(6522)
• SOUND VELOCITY PIPE(6524)
• LINER MATERIAL(6528)
If you edit these functions the pipe standard will be reset to the option OTHERS.
Factory setting
DIN PN10
NOMINAL DIAMETER
(6521)
!
Note!
This function does not appear if the option OTHERS was selected in the function
PIPE STANDARD (6520).
Use this function to select the nominal diameter of the pipe.
Options
OTHERS
DN:
25/1", 40/1½", 50/2", 80/3", 100/4", 150/6", 200/8", 250/10", 300/12",
400/16", 450/18", 500/20", 600/24", 700/28", 750/30", 800/32", 900/36",
1000/40", 1200/48", 1400/54", 1500/60", 1600/64", 1800/72", 2000/80"
!
Note!
The selection specifies the values for the following functions:
• CIRCUMFERENCE(6525)
• PIPE DIAMETER(6526)
• WALL THICKNESS(6527)
If you edit these functions the pipe standard will be reset to the option OTHERS and
the function NOMINAL DIAMETER (6521) does not appear.
Factory setting
80/3"