Description of Device Functions Proline Prosonic Flow 93T Portable
98 Endress+Hauser
11.7.4 Group DATA LOGGER
SAVE/LOAD/DELETE ?
(4993)
Use this function to save a site to the USB stick, load a site from the USB stick or delete
a site from the USB stick.
Options:
CANCEL
SAVE SITE
LOAD SITE
DELETE SITE
SAVE SITE stores the current programmed information to the USB stick.
!
Note!
Not all programmed information is saved. Only a subset of the total programmed
information is saved to a site. ä 46
LOAD SITE uploads to the programmed information from the site selected which is
stored on the USB stick to the instrument‘s ACTUAL SITE.
DELETE SITE removes the site from the USB stick.
!
Note!
It is recommended to delete sites from the USB stick via the Prosonic Flow 93T Site
Manager and not via PC or laptops. The name of the stored site is only visible via the
Site Manager.
Security Options:
NO
YES
Factory setting
NO
Functional description
OUTPUTS DATA LOGGER CONFIGURATION
LOG NAME
(4900)
Use this function to enter a log name for the data records (header) to be recorded.
!
Note!
After recording, the tag name is saved in the amplifier of the measuring device. The
name should be changed accordingly if the measuring point is changed.
User input
max. 8-character text, permitted characters are: A–Z, 0–9, +,–, punctuation marks
Factory setting
- - - - - - - - (No text)
LOGGING
(4901)
Use this function to start or stop logging data to the USB stick.
Options
OFF
ON
Factory setting
OFF
SYSTEM DATE/TIME
(4902)
Use this function to enter the current date and the current time.
User input
Day range: 00 to 99
Month range: 00 to 31
Year range: 00 to 12
Meridiem range A or P
LOGGING CYCLE
(4903)
Use this function to specify the intervals between the data logging cycles.
User input
Integers, 1 to 99999 s
Factory setting
10 s
Functional description
BLOCK OUTPUTS DATA LOGGER SITE MANAGER