Proline Prosonic Flow 93T Portable Description of Device Functions
Endress+Hauser 75
11.5.2 Group MAIN LINE
Function group CONFIGURATION
Functional description
USER INTERFACE MAIN LINE CONFIGURATION
1 = Main line
2 = Additional line
3 = Information line
A0001253
ASSIGN
(2200)
Use this function to define the display value assigned to the main line (the top line of
the local display) during normal measuring operation.
Options
OFF
VOLUME FLOW (CH1)
VOLUME FLOW IN % (CH1)
SOUND VELOCITY (CH1)
SIGNAL STRENGTH (CH1)
FLOW VELOCITY (CH1)
ACTUAL CURRENT 1
TOTALIZER (1 to 3)
ACTUAL CURRENT INPUT
VOLUME FLOW DEVICE 2
Factory setting
VOLUME FLOW DEVICES (CH1)
100% VALUE
(2201)
!
Note!
This function is not available unless VOLUME FLOW IN % was selected in the ASSIGN
function (2200).
Use this function to define the flow value to be shown on the display as the 100%
value.
User input
5-digit floating-point number
Factory setting
10 l/s
FORMAT
(2202)
Use this function to define the maximum number of places after the decimal point
displayed for the reading in the main line.
Options
XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX
Factory setting
X.XXXX
!
Note!
Note that this setting only affects the reading as it appears on the display, it has no
influence on the accuracy of the system's calculations.
The places after the decimal point as computed by the measuring device cannot
always be displayed, depending on this setting and the engineering unit. In such
instances an arrow appears on the display between the measuring value and the
engineering unit (e.g. 1.2 m3/h), indicating that the measuring system is
computing with more decimal places than can be shown on the display.