Description of Device Functions Proline Prosonic Flow 93T Portable
76 Endress+Hauser
Function group MULTIPLEX
Functional description
USER INTERFACE MAIN LINE MULTIPLEX
ASSIGN
(2220)
Use this function to define a second reading to be displayed in the main line alternately
(every 10 seconds) with the reading defined in the function ASSIGN (2200).
Options
OFF
VOLUME FLOW (CH1)
VOLUME FLOW IN % (CH1)
SOUND VELOCITY (CH1)
SIGNAL STRENGTH (CH1)
FLOW VELOCITY (CH1)
ACTUAL CURRENT 1
TOTALIZER (1 to 3)
ACTUAL CURRENT INPUT
VOLUME FLOW DEVICE 2
ACCUMULATOR CHARGE
Factory setting
OFF
100% VALUE
(2221)
!
Note!
This function is not available unless VOLUME FLOW IN % was selected in the ASSIGN
function (2220).
Use this function to define the flow value to be shown on the display as the 100%
value.
User input
5-digit floating-point number
Factory setting
Depends on nominal diameter and country, [value] / [dm
3
to m
3
or US-gal to US-Mgal]
Corresponds to the factory setting for the full scale value ( ä 118).
FORMAT
(2222)
Use this function to define the maximum number of places after the decimal point
displayed for the reading in the main line.
Options
XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX
Factory setting
X.XXXX
!
Note!
• Note that this setting only affects the reading as it appears on the display, it has no
influence on the accuracy of the system's calculations.
• The places after the decimal point as computed by the measuring device cannot
always be displayed, depending on this setting and the engineering unit. In such
instances an arrow appears on the display between the measuring value and the
engineering unit (e.g. 1.2 m3/h), indicating that the measuring system is
computing with more decimal places than can be shown on the display.