Description of Device Functions Proline Prosonic Flow 93T Portable
116 Endress+Hauser
11.10.2 Group VERSION INFO
Function group DEVICE
Function group SENSOR
SIMULATION MEASURAND
(8043)
!
Note!
This function is available in the SYSTEM group.
Use this function to set all inputs, outputs and totalizers to their defined flow-response
modes, in order to check whether they respond correctly. During this time, the words
"SIMULATION MEASURAND" appear on the display.
Options
OFF
VOLUME FLOW (CH1)
SOUND VELOCITY (CH1)
Factory setting
OFF
"
Caution!
• The measuring device cannot be used for measuring while this simulation is in
progress.
• The setting is not saved if the power supply fails.
VALUE SIMULATION
MEASURAND
(8044)
!
Note!
• This function is available in the SYSTEM group.
• The function is not visible unless the function SIMULATION MEASURAND (8043)
is active.
Use this function to specify a selectable value (e.g. 12 m
3
/s). This is used to test the
associated functions in the device itself and downstream signal loops.
User input
5-digit floating-point number, [unit]
Factory setting
0 [unit]
"
Caution!
• The setting is not saved if the power supply fails.
• The appropriate unit is taken from the function group SYSTEM UNITS (ACA)
( ä 69).
SYSTEM RESET
(8046)
Use this function to perform a reset of the measuring system.
Options
NO
RESTART SYSTEM (restart without interrupting power supply)
Factory setting
NO
Functional description
SUPERVISION VERSION INFO DEVICE
DEVICE SOFTWARE
(8100)
Displays the current device software version.
Functional description
SUPERVISION VERSION INFO SENSOR
SERIAL NUMBER
(8200)
Use this function to view the current serial number of the sensor.
Functional description
SUPERVISION SYSTEM OPERATION