Troubleshooting Proline Prosonic Flow 93T Portable
56 Endress+Hauser
9.5 Response of outputs to errors
!
Note!
The response of the totalizer (failsafe mode) can be configured by means of various functions in the
function matrix.
You can use positive zero return to set the signals to their fallback value, for example when
operation has to be interrupted while a pipe is being cleaned. This function has priority over all other
device functions; simulations are suppressed, for example.
9.6 Spare parts
Only accessories are available for the measuring device ä 49.
9.7 Return
ä 6
9.8 Disposal
Observe the regulations applicable in your country!
9.9 Software history
Error response mode of outputs and totalizers
Process/system error present Positive zero
return activated
"
Caution!
System or process errors defined as "notice messages" have no effect whatsoever on the inputs and outputs.
See the information on ä 38 ff.
Totalizer STOP
The totalizers stop if a fault is present.
ACTUAL VALUE
The fault is ignored. The totalizers continue to count in accordance with the
current flow measured value.
HOLD VALUE
The totalizers continue to count in accordance with the last valid flow measured
value (before the fault occurred).
The totalizer
stops.
Date Software version Changes to software Operating
Instructions
06.2011 2.03.XX – Added Site Manager
– Added Current Output
– Added Liner Information to Quick Set-up
– Added some ANSI standards to pipe standards
71136630/13.11
04.2010
2.02.XX No changes to software 71112144/04.10
06.2009 2.02.XX Original software for Prosonic Flow 93T Portable 71093720/06.09