Description of Device Functions Proline Prosonic Flow 93T Portable
68 Endress+Hauser
11.3 Block MEASURED VARIABLES
11.3.1 Group MEASURING VALUES´
Function group MAIN VALUES CH1
Functional description
MEASURED VARIABLES MEASURING VALUES MAIN VALUES CH1
The measuring values of channel 1 currently being measured are displayed in this function group.
!
Note!
The engineering units of all the measured variables shown here can be set in the SYSTEM UNITS group.
If the fluid in the pipe flows backwards, a negative sign prefixes the flow reading on the display.
VOLUME FLOW CH1
(0001)
The volume flow currently measured appears on the display (channel 1).
User interface
5-digit floating-point number, including unit and sign
e.g. 5.545 dm
/min; 1.4359 kg/h; 731.63 gal/d, etc.
SOUND VELOCITY
(0002)
The sound velocity currently measured in the liquid appears on the display (channel 1).
User interface
5-digit fixed-point number, incl. units
e.g. 1400.0 m/s, 5249.3 ft/s, etc.
FLOW VELOCITY CH1
(0003)
The flow velocity currently measured appears on the display (channel 1).
User interface
5-digit floating-point number, including unit and sign
e.g. 8.0000 m/s, 26.247 ft/s, etc.
SIGNAL STRENGTH CH1
(0007)
The signal strength appears on the display (channel 1).
User interface
4-digit fixed-point number
e.g. 80.0
!
Note!
To ensure reliable measurement takes place, Prosonic Flow requires a signal strength
of > 30.
VOLUME FLOW DEVICE 2
(0011)
The current actual value of the input current at the current input appears on the
display.
User interface
0.0 to 25 mA