Function group OPERATION
DEFINE PRIVATE CODE
(2021)
Use this function to enter a personal code number for enabling programming.
User input
0 to 9999 (max. 4-digit number)
Factory setting
93
!
Note!
• Programming is always enabled with the code "0".
• Programming has to be enabled before this code can be changed. When
programming is disabled this function is not available, thus preventing others from
accessing your personal code.
STATUS ACCESS
(2022)
Use this function to check the access status for the function matrix.
User interface
ACCESS CUSTOMER (parameterization possible)
LOCKED (parameterization disabled)
Factory setting
50%
ACCESS CODE COUNTER
(2023)
The number of times the private or service code was entered to access the device
appears on the display.
User interface
Integer (delivery status: 0)
Functional description
USER INTERFACE CONTROL OPERATION
TEST DISPLAY
(2040)
Use this function to test the operability of the local display and its pixels.
Options
YES
NO
Factory setting
OFF
Test sequence:
1. Start the test by selecting ON.
2. All pixels of the main line, additional line and information line are darkened for
minimum 0.75 seconds.
3. Main line, additional line and information line show an "8" in each field for
minimum 0.75 seconds.
4. Main line, additional line and information line show a "0" in each field for
minimum 0.75 seconds.
5. Main line, additional line and information line show nothing (blank display) for
minimum 0.75 seconds.
When the test completes the local display returns to its initial state and the setting
changes to OFF.
Functional description
USER INTERFACE CONTROL UNLOCKING/LOCKING