Proline Prosonic Flow 93T Portable Description of Device Functions
Endress+Hauser 109
VISCOSITY
(6543)
This function displays the viscosity of the liquid. This is determined via the values
entered in the LIQUID (6540) and TEMPERATURE (6541) functions.
If you edit the predetermined value the function LIQUID (6540) will be set to the
option OTHERS.
The viscosity has to be entered if the liquid is not available for selection in the function
LIQUID (6540) and the OTHERS option was selected.
User input
Fixed-point number 0.0 to 5000.0 mm
2
/s
Factory setting
l mm
2
/s
SOUND VELOCITY
NEGATIVE
(6545)
Use this function to specify the lower search range for the sound velocity of the liquid.
User input
Fixed-point number 0 to 1000 m/s
Factory setting
500 m/s
!
Note!
Please refer to the explanations in the function SOUND VELOCITY LIQUID (6542).
SOUND VELOCITY POSITIVE
(6546)
Use this function to specify the upper search range for the sound velocity of the liquid.
User input
Fixed-point number 0 to 1000 m/s
Factory setting
300 m/s
!
Note!
Please refer to the explanations in the function SOUND VELOCITY LIQUID (6542).