Installation Proline Prosonic Flow 93T Portable
14 Endress+Hauser
Running the "Sensor Installation" Quick Setup menu
!
Note!
• If you are not familiar with the operation of the device ä 33.
• The following section only describes the steps necessary for Clamp On type of mounting within
the "Sensor Installation" Quick Setup.
Running the Quick Setup for Clamp On type of mounting
1. Enter or select installation-specific values or the values specified here.
2. Read off the installation distances necessary for mounting.
Subsequent procedure
The sensors can be installed once the installation distances have been determined:
• Prosonic Flow P (DN 15 to 65 / ½ to 2½") ä 23
• Prosonic Flow P (DN 50 to 4000 / 2 to 160") ä 23
Home position Quick Setup Setup sensor
Language
Clamp On Measurement
Sensor type
1 or 2 Sensor configuration = Number of traverses
Pipe standard
Nominal diameter
Pipe material
Sound velocity pipe
Pipe diameter
Circumference
Wall thickness
Liner material
Sound velocity, Liner
Liner, thickness
Liquid
Installation
distances for
measurement via
one traverse:
Installation
distances for
measurement via
two traverses:
Temperature
Sound velocity liquid
Position sensor –
Wire length –
Sensor distance
No Other measurement?
Setup sensor