EasyManua.ls Logo

Endress+Hauser Proline Prosonic Flow 93T Portable - Page 27

Endress+Hauser Proline Prosonic Flow 93T Portable
124 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Prosonic Flow 93T Portable Installation
Endress+Hauser 27
A0001117
Fig. 28: Mounting the sensor holders
5. Coat the contact surfaces of the sensors with an even layer of coupling fluid
approx. 1 mm (0.04") thick, going from the groove through the center to the opposite edge.
A0011373
Fig. 29: Coating the contact surfaces of the sensor with coupling fluid
6. Insert the sensor into the sensor holder.
7. Fit the sensor cover on the sensor holder and turn until:
The sensor cover engages with a click.
The arrows (Å / Æ "close") are pointing towards one another.
8. Screw the connecting cable into the individual sensor.
A0011376
Fig. 30: Mounting the sensor and connecting the connecting cable
This completes the mounting process. The sensors can now be connected to the transmitter via the
connecting cables ä 31.

Table of Contents