Proline Prosonic Flow 93T Portable Commissioning
Endress+Hauser 47
All other programming data in the instrument's memory remains unchanged when sites are saved
or loaded. This means that other parameters not included in the list above need to be managed
manually via the keypad and program menu.
!
Note!
The Correction Factor (BASIC FUNCTIONS F SENSOR DATA F CALIBRATION DATA) is
set to 1.00 for all sites saved to the USB stick. If correction factors are applied in the field these
values need to be manually reset via the keypad and program menu. A Correction Factor is valid
until the site is saved or a new site is loaded.
!
Note!
The Zero Point (BASIC FUNCTIONS F SENSOR DATA F CALIBRATION DATA) is set to
0.000 nanoseconds for all sites saved to the USB stick. If a zero point adjustment is made in the field
these saves need to be manually reset via the keypad and program menu. A zero point adjustment
is valid until teh site is saved or a new site is loaded.
!
Note!
The Arbitrary Volume Function is not supported by the Site Manager. No custom volume units will
be stored in the saved site information.
!
Note!
All Site Manager functions are disabled if a data log is running.
!
Note!
It is recommended to delete sites from the USB stick via the Prosonic Flow 93T Site Manager and
not via PC or laptops. The name of the stored site is only visible via the Site Manager.
Group LIQUID DATA LIQUID (6540)
TEMPERATURE (6541)
SOUND VELOCTITY LIQUID (6542)
VISCOSITY (6543)
Group TOTALIZER (1..3) ASSIGN (3000)
UNIT TOTALIZER (3001)
TOTALIZER MODE (3002)