Proline Prosonic Flow 93T Portable Troubleshooting
Endress+Hauser 53
No. # 6xx Simulation operation active
601 S: POS. ZERO RET.
!: # 601
Positive zero return active.
"
Caution!
This notice message has the highest
display priority!
Switch off positive zero return.
661
to
664
S: SIM. CURR. INP.
!: # 661 to 664
Current input simulation active. Switch off simulation.
691 S: SIM. FAILSAFE
!: # 691
Simulation of failsafe mode (outputs)
active.
Switch off simulation.
692 S: SIM. MEASURAND
!: # 692
Simulation of a measured variable active
(e.g. mass flow).
Switch off simulation.
698 S: DEV. TEST ACT.
!: # 698
The measuring device is being checked on
site via the test and simulation device.
743
to
745
S: 0-ADJ.FAIL CHn
!: # 743 to 745
The static zero point calibration of
Channel 1/2 is not possible or was
interrupted.
Check that the flow velocity is = 0 m/s.
752 S: W. THICKNESS CH 1
!: # 752
Channel 1: Wall thickness measurement
active
Switch off wall thickness measurement
No. Error message/type Cause Remedy (spare parts ä 49 ff.)