Proline Prowirl 72 FOUNDATION Fieldbus 11 Operation via FOUNDATION Fieldbus
Endress + Hauser 105
Totalizer -
Assign
AUTO - OOS Use this function to assign a measured variable for the totalizer.
Options:
Off
Volume Flow
Calc. Mass Flow
Corr. Volume Flow
Factory setting:
Volume Flow
! Note!
The totalizer is reset to "0" as soon as the selection is changed.
Totalizer -
Reset
AUTO - OOS Use this function to reset the totalizer ("Totalizer – System Value"
parameter) to 0 (= RESET).
Options:
No
Yes
Factory setting:
No
! Note!
Totalizer resetting can be controlled or triggered via FOUNDATION
Fieldbus by means of cyclic data transfer via the Discrete Output Function
Block (Page 119 ff.).
Totalizer -
Reset All
AUTO - OOS Use this function to reset the totalizer ("Totalizer – System Value"
parameter) to 0 (= RESET).
Options:
No
Yes
Factory setting:
No
! Note!
Totalizer resetting can be controlled or triggered via FOUNDATION
Fieldbus by means of cyclic data transfer via the Discrete Output Function
Block (Page 119 ff.).
Transducer Block "Totalizer" (Endress+Hauser parameters) / base index 600
Parameter Write access for
operating mode
(MODE_BLK)
Description