SYSTEM PARAMETERS
POS. ZERO
RETURN
(p. 77)
OFF
–ON
DEF. PRIVATE
CODE
(p. 77)
max. 4-digit number: 0...9999
Factory setting:
33
Your setting: ............................................
ACCESS CODE
(p. 78)
max. 4-digit number: 0...9999
Factory setting:
0
SELF CHECKING
(p. 78)
OFF –
ON
Your setting: ............................................
PRESENT SYSTEM
CONDITION
(p. 79)
Display (entries chronological):
F: ... = Error message (system error)
A: ... = Alarm message (process error)
S: ... = Status message
S: SYSTEM WORKS NORMALLY
PREVIOUS
SYSTEM
CONDITIONS
(p. 79)
Display (chronological, max. 10 entries):
F: ... = Error message (system error)
A: ... = Alarm message (process error)
S: ... = Status message
S: NO ENTRY EXISTING
SOFTWARE
VERSION
(p. 80)
Display:
e.g. PRO 33 V3.01.00
SOFTWARE VER.
COM
(p. 80)
Display:
e.g. V3.02.00 HART; V3.02.00 RS 485
SENSOR DATA
Normally these characteristic sensor data may not be altered.
A change to the sensor data affects a number of functions of the
whole measuring system, especially its accuracy! The following
functions of this group can therefore only be changed after
entering a
service code
and cannot be altered using the
personal code.
Please contact your E+H Service organisation for more
information.
K-FACTOR
POS.
(p. 81)
max. 5-digit fixed point:
0.5000...2.0000
Factory setting:
dependent
on the sensor
(DN) and its calibration
Setting: ...................................................
K-FACTOR
NEG.
(p. 81)
max. 5-digit fixed point:
0.5000...2.0000
Factory setting:
dependent
on the sensor
(DN) and its calibration
Setting: ...................................................
SENSOR DATA (continued)
ZERO POINT
(p. 81)
max. 4-digit number: –1000...+1000
Factory setting:
dependent
on the sensor
(DN) and its calibration
Setting: ..................................................
NOMINAL
DIAMETER
(p. 81)
Selected from fixed table:
2...2000 mm or 1/12...78 inch
Factory setting:
dependent
on the sensor (DN)
Setting: ..................................................
MAX. SAMPLING
RATE
(p. 82)
max. 3-digit fixed point:
1.0...60.0 / s (per second)
Factory setting:
dependent
on the sensor (DN)
Setting: ..................................................
SAMPLING RATE
(p. 82)
max. 3-digit fixed point: 1.0...60.0 /s
(upper limit as for MAX. SAMPLING RATE)
Factory setting:
dependent
on the sensor (DN)
Setting: ..................................................
SERIAL NUMBER
(p. 82)
max. 6-digit number: 1...999999
Setting: ..................................................
EPD ELECTRODE
(p. 82)
YES *
–NO
* with EPD electrode as standard
Setting: ..................................................
POLARITY ECC
(p. 83)
POSITIVE – NEGATIVE
Factory setting (Electrode material):
Positive → 1.4435, Hastelloy-C, Platinum
Negative → Tantalum
Setting: ..................................................
10 Functions at a Glance Promag 33
120 Endress+Hauser