EasyManuals Logo

ESAB fabricator 141i Service Manual

ESAB fabricator 141i
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
ESAB FABRICATOR 141i
Manual 0-5448 6-19 TROUBLESHOOTING
6.13 Calibration
1 Calibration
Imin
Vmin
Imax
W2
WVF
A
SW0
Art # A-12224
Set SW0-1 to ON, SW0-2 to ON, SW0-3 to ON, SW0-4 to OFF to allow calibration of output volts & amps, and calibration of wire feed
speed.
2 Output Current Calibration
Select STICK mode on the front panel.
Measure open circuit voltage and adjust WVF potentiometer so Volts display reads within 0.2V of the measured value.
Connect a load to the output terminals. The load should be of a resistance to give 25V at 250A.
Set front panel AMPS potentiometer to minimum.
Adjust Imin trimpot until output amps is 10A +/- 0.2A
Set front panel AMPS potentiometer to maximum.
Adjust Imax trimpot until output amps is 90A +/- 1A
Recheck settings
Set front panel AMPS potentiometer to maximum.
Adjust A potentiometer so Amps display reads within 0.5A of the measured value.
3 Output Voltage Calibration
Select MIG mode on the front panel.
Remove the load from the output terminals.
Set front panel VOLTS potentiometer to minimum.
Adjust Vmin trimpot until output volts is 10.0V +/- 0.2V
Set front panel VOLTS potentiometer to maximum.
Adjust W2 trimpot until output volts is 19V +/- 0.2V
Recheck settings
4 Wire Speed Calibration
NOTE!
These adjustments are on the wiring side of the Main PCB1.

Table of Contents

Other manuals for ESAB fabricator 141i

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ESAB fabricator 141i and is the answer not in the manual?

ESAB fabricator 141i Specifications

General IconGeneral
BrandESAB
Modelfabricator 141i
CategoryWelding System
LanguageEnglish

Related product manuals