96
8. Rinse the probe with de-ionized water. Place it in the next solution and
follow the steps 2 to 7 to calibrate other points, if ‘MULTI’ point calibration is
selected.
9. Press ESC (F4) to go to measurement mode. The meter shows the
resistivity reading of the solution in measurement mode.
2.1 Resistivity Calibration Report
Calibration report gives you information on calibration. It includes date &
time of calibration and the cal factor.
2.1.1 To View Calibration Report:
1. From resistivity measurement mode, press left or right arrow key to
navigate to other available functions until you see REPO function in the LCD
2. Press REPO (F2) key. The calibration report is shown in the display
3. Press PRIN (F2) to transfer the calibration report to the computer.( Refer
page Error! Bookmark not defined. to print data)
Figure 47 : Resistivity Report Screen
Note: The function of the keys available on the resistivity cell constant/
calibration/ report screen is the same as previously described in
conductivit
y
calibration.