60
1. Ion Setup
Ion setup sub-group allows you to configure Ion measurement & calibration related
settings of the meter.
Ion Settings Page
Figure 23: Ion Settings Page
Parameter Description Factory Default
Ion unit
Sets the unit of measurement for Ion.
Available units: ppm, molar and mg/L
ppm
) Press (Up) or (Down) arrow key to change the value of a selected parameter.
2. Ion Calibration Mode
2.1 About Ion Calibration
The meter supports Ion calibration up to 8 points with minimum of 2-points. The
available 8-points are 0.001, 0.01, 0.1, 1, 10, 100, 1000 & 10000.
The meter guides your through the calibration process with on-screen
instructions and hints. If the second point and subsequent points are not within
the allowable calibration window (15mV/decade to 90mV/decade), the meter will
not accept the calibration.
When you try to re-calibrate the meter with ISE, the meter gives you 2 options:
Clear the previous calibration (choose this if you wish to delete all
previously calibrated points and re-calibrate with a new type of Ion)
Retain the previous calibration (choose this if you wish to retain
previously calibrated points and re-calibrate new points or overwrite
existing points with same type of Ion).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EUTECH INSTRUMENTS pH 600 and is the answer not in the manual?

EUTECH INSTRUMENTS pH 600 Specifications

General IconGeneral
BrandEUTECH INSTRUMENTS
ModelpH 600
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals