112
which the meter is going to be calibrated. Press & keys to adjust
the reading to the known oxygen concentration of the sample.
7. Press ENTER key to confirm the entered value.
8. Press ESC (F4) to go to measurement mode. The meter shows the
concentration reading of the solution in measurement mode.
Figure 59 : Concentration Calibration
2.1 Concentration Calibration Report
Calibration report gives you information on calibrated temperature and
calibration at pressure and salinity. It includes date & time on which the last
calibration was done and the offset.
2.1.1 To View Calibration Report:
1. From concentration measurement mode, press left or right arrow
key to navigate to other available functions until you see REPO
function in the LCD.
2. Press REPO (F2) key. The calibration report is shown in the display
Press PRIN (F2) to transfer the calibration report to the computer.
Note: The function of the keys available on the concentration calibration/
re
ort screen is same as
reviousl
described in saturation calibration.