Fiber Inspection Scope 93
8 Troubleshooting
Solving Common Problems
Problem Possible Cause Solution
My unit does not start. The battery is
completely discharged
(if the battery level
allows it, the battery LED
will remain red for about
5 seconds when you try
to turn on the unit).
Connect the unit to an external
power source to recharge the
battery. If the battery is no longer
charging properly, you may need to
replace it with a new one
(see Replacing the Battery on
page 84).
The system has
encountered a problem.
Press the on/off button for at least
ten seconds to force a hardware
reset on the unit.
a
Some files essential to
the normal operation of
the unit have been
corrupted.
Press the on/off button for at least
ten seconds to force a hardware
reset on the unit.
a
If the problem persists, try resetting
the FIP-500 to its factory settings
(see Reverting to Factory Settings
on page 38).
My unit is not
responding.
The system has
encountered a problem.
Press the on/off button for at least
ten seconds to force a hardware
reset on the unit.