A1
Basic user manual
Finapres Nova
53
Item
Description
NIBP FINGER STOP
Stops the finger blood pressure measurement. This action does
not affect the state of recording.
NIBP ARM START
Starts the upper arm blood pressure measurement.
NIBP ARM STOP
Stops the upper arm blood pressure measurement.
MEASUREMENT SETTINGS
Opens the MEASUREMENT SETTINGS screen that allows the
user to review and change the user bound measurement
settings.
If the finger cuff needs to be re-adjusted during a recording, the
button can be used to pause the finger blood pressure measurement
without interrupting the recording of the other signals. After re-
adjustment the finger blood pressure measurement can be restarted
using button
If an error occurs which pauses the finger blood pressure measurement,
it can be restarted using the button without stopping the recording.
In case the measurement is prolonged (e.g. due to patient movement) it
is the user’s responsibility to verify that the upper arm measurement is
finished successful or manually aborted within acceptable time.
An upper arm measurement on the contralateral arm is less accurate.