A2
Basic user manual
Finapres Nova
86
A2.2.3 Cuff pressure
A cuff pressure greater than 250 mmHg sustained for 2.5 s depressurizes the
cuff within 1 second.
During the start procedure cuff pressure is limited to a maximum of
295 mmHg lasting less than 2 s.
A2.2.4 General
If cuff pressure oscillates when measuring finger pressure, although this
presents no hazard or discomfort to the patient, software takes action to
remove the oscillation.
Fully contracted finger arteries do not allow pressure monitoring. A warning is
send to the Finapres Nova and the finger cuff is depressurized when this is
detected. The Finapres Nova waits maximally 100 s and starts up again.
A2.3 Signals
The table below gives an overview of the available signals.
Abbreviation
Name
description
min
max
unit
reBAP
Reconstructed
Brachial Arterial
Pressure
reconstructed
brachial pressure
from finger pressure
waveform
0
300
[mmHg]
reSYS
reBAP Systolic
Pressure
systolic brachial
pressure (based on
reBAP)
0
300
[mmHg]
reDIA
reBAP Diastolic
Pressure
diastolic brachial
pressure (based on
reBAP)
0
300
[mmHg]
reMAP
Mean arterial
pressure
mean arterial
pressure (based on
reBAP)
0
300
[mmHg]
fiAP
Finger Arterial
Pressure
finger arterial
pressure
0
400
[mmHg]
fiSYS
FinAP Systolic
Pressure
SYStolic pressure
(from BP signal)
0
400
[mmHg]
fiDIA
FinAP Diastolic
Pressure
DIAstolic pressure
(from BP signal)
0
400
[mmHg]
fiMAP
FinAP Mean
Arterial
Pressure
Mean Arterial
Pressure (from BP
signal)
0
400
[mmHg]
HR
Heart rate
Pulse Rate derived
from best signal
0
300
[beats/
min]
HR AP
AP Heart rate
Pulse Rate derived
from arterial
pressure signal
0
300
[beats/
min]
IBI
Inter Beat
Inter Beat Interval
0
2000
[ms]