Version date: 04 April 2016
A2.4.5 EMC Guidance and manufacturer's declaration tables
This section describes the information regarding potential electromagnetic or
other interference and advice regarding avoidance.
Guidance and manufacturer's declaration - electromagnetic emissions
The Finapres Nova is intended for use in the electromagnetic environment
specified below. The customer or the user of the Finapres Nova should assure that
it is used in such an environment.
Electromagnetic environment -
guidance
The Finapres Nova uses RF energy only
for its internal function. Therefore, its
RF emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
Harmonic
emissions
IEC 61000-3-2
Not applicable
(< 75 W power)
The Finapres Nova is suitable for use in
all establishments including domestic
and those directly connected to the
public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Voltage fluctuations
/ flicker emissions
IEC 61000-3-3
Not applicable
(low
power/limiting
switching)
Guidance and manufacturer's declaration - electromagnetic immunity
The Finapres Nova is intended for use in the electromagnetic environment
specified below. The customer or the user of the Finapres Nova should assure
that it is used in such an environment.
Electromagnetic
environment - guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for
power supply
lines
±1 kV for
input/output
lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.