7 Installation of the Transducers FLUXUS WD100, WD200
UMFLUXUS_F5WDV1-1EN, 2016-04-08 35
7.3.2 Installation of the Rail
Select the installation instructions that correspond to the supplied clasp:
see section Installation of the rail without a clasp
see section Installation of the rail with the ratchet clasp
Installation of the rail without a clasp
Cut the tension strap to length (pipe circumference + at least 120 mm).
Insert approx. 100 mm of the tension strap into one of the slots of the tension strap clamp and bend it (see Fig. 7.7).
If necessary, insert the long end of the tension strap into the metal spring (see Fig. 7.8). It is not necessary to mount the
metal spring:
on steel pipes
on pipes with an outer pipe diameter < 80 mm or
on pipes that are not subjected to significant temperature fluctuations
Place the tension strap around the pipe (see Fig. 7.9).
Position the metal spring (if mounted) and the tension strap clamp (see Fig. 7.9):
On a horizontal pipe, mount the tension strap clamp on the side of the pipe, if possible.
Mount the metal spring (if necessary) on the opposite side of the tension strap clamp.
Fig. 7.7: Tension strap with tension strap clamp
Fig. 7.8: Tension strap with the metal spring and the tension strap clamp
Fig. 7.9: Tension strap with the metal spring and the tension strap clamp on the pipe
tension strap clamp
tension strap clamp
metal spring
metal spring
s
ab
tension
strap
metal spring
s = length of the rail - 33 mm