FLUXUS WW 16 Outputs
120 UMFLUXUS_WWV1-1EN, 2018-02-28
When adjusting to changed measurement conditions e.g, a substantial increase of the fluid temperature, the alarm will not
be switched. Alarm outputs with the switching condition OFF will be set automatically to the switching function NO Cont..
16.6.5 Alarm state indication
The alarm state can be displayed after the configuration of the alarm outputs and during the measurement. This function is
activated in Special Funct.\SYSTEM settings\Dialogs/Menus. The activation of this function is recommended
when alarm outputs have to be reconfigured frequently.
If the alarm state indication is activated, the alarm output state is displayed after configuring the alarm outputs:
Note! There is neither a visual nor an acoustic indication of the alarm output switching.
Select the menu item SHOW RELAIS STAT. Select on to activate the alarm state indica-
tion.
The alarm state indication is structured as follows:
Rx = , with X being the number of the alarm output and a picto-
gram according to Tab. 16.6.
The configuration of the alarm outputs can be repeated by pressing key C. When the con-
figuration of the alarm outputs is finished, press ENTER. The main menu will be displayed.
If the alarm output indication is activated, the alarm state can be displayed during the measurement. Press key to
scroll through the upper line or key in the lower line until the alarm state is displayed.
Tab. 16.6: Pictograms for the alarm state indication
No. func
(switching condition)
typ
(holding behavior)
mode
(switching function)
current
state
R=
1 OFF NON-HOLD NO Cont. CLOSED
2 MAX HOLD NC Cont. OPEN
3 MIN
+- -+
QUANT.
ERROR
Example: R1 =
SHOW RELAIS STAT
off >ON<
R1= R2=
R3= C=REPEAT