13 Data logger and transmission of data FLUXUS WW
UMFLUXUS_WWV1-1EN, 2018-02-28 95
Max. 100 series of measured values can be stored. The number of series of measured values depends on the total num-
ber of measured values stored in the previous series of measured values.
It is possible to display the time at which the data logger will be full during the measurement. All activated channels, total-
izers and other values will be considered.
13.2 Data transmission
The measured values can be transmitted to a PC via the serial interface RS232 or RS485 (optional) or stored on a SD
card (optional).
13.2.1 Online transmission
The measured values are transmitted during the measurement. The measurement data can be transmitted to a terminal
program.
The data logger works independently of the online transmission.
13.2.2 Offline transmission
The measurement data of the data logger are transmitted.
Selection of the serial interface for the offline transmission
Select Special Funct.\SYSTEM settings\serial transmis. Press ENTER until Send Offline via is dis-
played.
Press key to scroll through the display of the upper line.
If the ringbuffer is activated and has overflown at least once, this display will be indicated.
Tab. 13.1: Overview of online transmission
serial interface transmission see
RS232 terminal program section 13.2.5
RS485 (sender) terminal program section 13.2.5
Note! It is recommended to use the RS485 interface for the online trans-mission of data. The RS232 inter-
face should only be used if the transmitter does not have an RS485 interface.
Tab. 13.2: Overview of offline transmission
serial interface transmission see
RS232 terminal program section 13.2.6
RS485 (sender) terminal program section 13.2.6
Select the serial interface for the offline transmission of da-ta.
This display will only be indicated if the transmitter has an RS485 interface.
full= 26.01/07:39
54.5 m3/h
last= 26.01/07:39
54.5 m3/h
Send Offline via
RS232 >RS485<