FLUXUS WW 6 Measuring point selection
26 UMFLUXUS_WWV1-1EN, 2018-02-28
disturbance source: reducer
inlet: l ≥ 10 D outlet: l ≥ 5 D
disturbance source: pump
inlet: l ≥ 20 D
Tab. 6.2: Recommended distance from disturbance sources; D – nominal pipe diameter at the measuring point, l – recommended distance