8 Start-up
PIOX R721 8.5 Initialization
49
UMPIOX_F72xV1-3EN, 2021-09-01
8.5 Initialization
During an initialization (INIT) of the transmitter, all settings are reset to the factory settings.
Proceed as follows to execute an initialization:
• When switching on a transmitter: keep and C pressed.
• During the operation of the transmitter: press , C and ENTER simultaneously. Release only ENTER. Keep
and C pressed.
During an initialization it is tested whether the key lock is activated. If so, it has to be deactivated.
• Enter a 6-digit key lock code.
• Press ENTER.
If a measurement is currently running, it will be stopped.
It will be requested whether the initial settings are to be carried out.
If Yes is selected, the following setting dialogs will be displayed.
•Language
•Date/time
•Units of measurement
•Delete meas. values
• Delete user subst. (all customized fluids which were stored after delivery will be deleted)
The initialization can also be started with the HotCode 909000.
8.6 Date and time
The transmitter has a battery-powered clock. Measured values are automatically stored with date and time.
• Select the menu item Date/time.
The adjusted time is displayed.
• Enter the current time via the numeric field.
• Press ENTER.
The adjusted date is displayed.
• Enter the current date via the numeric field.
• Press ENTER.
Initial settings
Miscellaneous\System settings\Date/time