FT-107 Weighing Terminal, Technical Manual, Rev. 1.0.0, August 2019 Page 37 of 46
7.3 Fast Access to the Calibration
The instrument has fast access calibration feature to save time for the service technician. If only the calibration
adjustment is needed, follow the steps below for fast calibration.
Display
Operation
[123.456 ] kg
Press STP key until [ PASS ] prompt appears and
confirm with Enter (only non-OIML scales ).
Or press calibration switch.
[ PASS ]
Enter “011012” with navigation keys.
[ 011012 ]
Press Enter key for confirm password.
[ bUıLd ]
First block of set-up. Press Zero key longer than 2 seconds to
jump to the zero adjustment until the prompt on the below seen.
[ CAL ]
Zero adjustment: Unload the scale and press Enter key. Do not
touch the scale at [ - - - ] prompt.
[ LoAd ]
[ XXXXX ]
Span adjustment: Load the scale and enter the load value by
pressing navigation keys and press Enter key. Do not touch the
scale at [ - - - ] prompt.
[ LınCAL ]
The next item Linearity Correction prompt appears after the
calibration.