FT-107 Weighing Terminal, Technical Manual, Rev. 1.0.0, August 2019 Page 42 of 46
[ SPA.AdJ ] Span Adjustment
This parameter lets you to perform span adjustment.
If you will not perform Span adjustment press ► key or press Esc key to return to the main block.
Span adjustment is performed as;
1. Press Enter key to start the span adjustment.
2. After a while the [ LoAd ] prompt appears with a suggested test weight on the display as
[ XXXXXX ]. For definition of different test weight use the navigation keys.
FLINTEC recommends to use between 50 and 100% of the scale capacity but minimum 20%.
An insufficient weight leads to a calibration error.
3. Place the test weights or another practical weight on the scale.
4. Press Enter to start span adjustment. [ - - - ] message will be shown on the display while span
calibration is being performed.
5. The next item Span Adjustment Under Load prompt appears after the calibration.
[ SUL.AdJ ] Span Adjustment Under Load
This parameter is being used to perform span adjustment of a scale without lifting the load. This
operation is especially recommended for span adjustment for filled tanks without emptying the
material.
If you will not perform Span Adjustment Under Load press ► key or press Esc key to return to
the main block.
Span Adjustment Under Load is performed as;
1. Press Enter key to start the span adjustment under load.
2. At the [ UnLoAd ] prompt, press Enter for temporary zero adjustment.
3. After a while the [ LoAd ] prompt appears with a suggested test weight on the display as
[ XXXXXX ]. For definition of different test weight use the navigation keys.
FLINTEC recommends to use between 50 and 100% of the scale capacity but minimum 20%.
An insufficient weight leads to a calibration error.
4. Place the test weights or another practical weight on the scale.
5. Press Enter to start span adjustment. [ - - - ] message will be shown on the display while span
calibration is being performed.
6. The next item appears after the calibration.
7.5.5 [ GrAvit ] Gravity Adjustment
For gravity adjustment before shipment of the scale, enter the values below.
If you will not adjust the gravity press ► key to enter the next item.
1. Press Enter key.
2. [ CAL-Gr ] Gravity of the manufacturer. Press Enter key and enter the gravity coefficient of the
place of the scale calibration via navigation keys. Press Enter key.
3. [ USAGE ] Gravity of the usage. Press Enter key and enter the gravity coefficient of the place of
the scale usage via navigation keys.
Warning: Check the scale accuracy after gravity adjustment against wrong entry.