Option -01 Battery Pack
Calibration
7
7-11
3. With no ac line power applied to the meter, adjust the variable voltage power supply
for 8.5 V dc output.
4. Now apply ac line power, and check that the charging current is greater than 250
mA.
5. Slowly increase the variable voltage supply output. Check for the following:
The charging current decreases as the supply voltage approaches 10 V dc.
The current should continue to decrease as the voltage increases, until the
current is between 60 and 80 mA.
With any additional decrease in current (increase of the supply voltage), the
current should suddenly jump to a low value (less than 0.5 mA.)
When the variable voltage supply output is reduced to 9.35 V dc, the current
should be between 5 and 15 mA.
BATTERY OPTION PCA
RED
BLACK
IV
+
-
POWER SUPPLY
0 TO 15V
OPTIONAL VOLTMETER
TO MONITOR A BATTERY VOLTAGE
METER TO MEASURE
CHARGING CURRENT
33
5
.1 SHUNT
qb36f.eps
Figure 7-5. Cycle/Float Charge Rate Switch Test