Option -05 IEEE-488 Interface
Performance Testing
8
8-9
5. Using needle nose pliers, detach the two ribbon cables at the front part of the IEEE-
488 PCA. Alternately pull up on each end of the cable connectors.
6. Remove the IEEE-488 PCA, disengaging the assembly from both the small slot in
the side of the meter and the plastic standoff at the front corner of the assembly.
8-12. Installing the IEEE-488 Interface Option
Use the following procedure to install the IEEE-488 Interface Option.
1. Turn the meter off, and unplug the power connection.
2. Check that the plastic standoff remains in the appropriate hole in the meter PCA
(narrow end of the standoff down; see Figure 8-3).
3. If necessary, install the two ribbon cables on the IEEE-488 PCA. Each cable fits in
only one socket and in only one direction. Make sure the cables lock firmly in place.
4. Attach opposite ends of the ribbon cables onto the Fluke 45 Main PCA.
5. Install the IEEE-488 PCA in the Fluke 45 with the dual ribbon cables facing the
front of the meter. The IEEE-488 PCA slips into the small slot in the side of the
meter. The end of the plastic standoff fits into the hole in the IEEE-488 PCA. Make
sure the IEEE-488 PCA is firmly gripped against the retainer on the standoff. The
rear of the IEEE-488 PCA should rest upon the support just forward of the
transformer.
6. Secure the rear of the IEEE-488 PCA with the panhead Phillips screw.
7. Connect the 24-line cable assembly to the IEEE-488 PCA. (See Figure 8-2.)
8. Reinstall the meter case so it seats properly in the front panel. Attach the rear bezel
with the two panhead Phillips screws, and secure the case with the flathead Phillips
screw in the bottom.
8-13. Performance Testing
Use the performance test program in Figure 8-1 to verify operation of the IEEE-488
Interface. This program is written for use with the Fluke 182A Instrument Controller and
its interpreted BASIC language. The program may be adapted to the language of any
IEEE-488 controller.
This performance test communicates to a meter that has been configured for IEEE-488
operation at address 0. Lines 160 and 170 initialize the IEEE-488 bus and send a
selective device clear to the meter. A multiple byte command is sent to the meter (by line
190) to clear the meter status. Another command sequence (including a query) is sent to
the meter by line 210; the meter asserts Service Request (SRQ) to signal that a response
is available. Lines 530 through 560 first poll the meter for status, then input the response
from the meter. Lines 230 through 270 test for proper operation and print the results.