Performance Testing and Calibration
Introduction
4
4-3
4-1. Introduction
This chapter of the Service Manual provides performance tests that can be used at any
time to verify Fluke 45 operation within published specifications. A complete calibration
procedure is also included. The performance test and, if necessary, the calibration
procedure can be performed periodically and after service or repair.
4-2. Required Equipment
Equipment required for performance testing and calibration is listed in Table 4-1.
Table 4-1. Recommended Test Equipment
Instrument
Type
Minimum Specifications
Recommended
Model
Multifunction
Calibrator
DC Voltage:
Range = 90 mV to 1000 V dc
Accuracy = .005%
Fluke 5700A
with wideband
option
AC Voltage:
Frequency Voltage Accuracy
1 kHz 29 mV to 750 V 0.05%
100 kHz 15 mV to 300 mV 1.25%
AC Milliamps:
Frequency Current Accuracy
1 kHz 29 mA to 100 mA 0.125%
DC Milliamps:
Frequency Voltage Accuracy
10 kHz 1 V 0.125%
Boost
Amplifier
DC Amps: Current
10 A
Accuracy
.05%
Fluke 5725A
AC Amps:
1 kHz 2A and 10 A 0.25%
Decade
Resistance
Source
Ohms Accuracy
General
Resistance Inc.
Model RDS 66A
290 0.0125%
2.9 k 0.0125%
29 k 0.0125%
290 k 0.0125%
2.9 M 0.0125%
Note
The 5700A Calibrator can be used for 0.05% accuracy (rated) on the
3.0 k
, 30 k
, 300 k
, and 3.0 M
ranges. This calibrator can be used
for 0.06% accuracy on the 100
and 300
ranges.