Introduction and Specifications
Specifications
1
1-11
Maximum Crest Factor
3.0
Maximum Input
mA 300 mA dc or ac rms. Protected with a 500 mA, 250 V, IEC 127-sheet I, fast
blow fuse and a 440 mA, 1000 V, fast blow fuse.
A 10 A dc or ac rms continuous, or 20 A dc or ac rms for 30 seconds maximum.
Protected with a 11 A, 1000 V, 17,000 A interrupt rating, fast blow fuse.
Note
Resistance between the COM binding post and the meter’s internal
measuring circuits is approximately .003
.
Ohms
Range Resolution Accuracy
Typical
Full Scale
Voltage
Max Current
Through the
Unknown
Slow Medium Fast
300 - 10 m 100
m
0.05% + 2 + 0.02 0.25 1 mA
3 k - 100 m 1 0.05% + 2 0.24 120 µA
30 k -1 10 0.05% + 2 0.29 14 µA
300 k - 10 100 0.05% + 2 0.29 1.5 µA
3 M - 100 1 k 0.06% + 2 0.3 150 µA
30 M -1 k 2% 0.25% + 3 2.25 320 µA
300
M*
- 100 k 1 M 2% 2.0 320 µA
100 1 m - - 0.05% + 8 + 0.02 0.09 1 mA
1000 10 m - - 0.05% + 8 + 0.02 0.10 12 µA
10 k 100 m - - 0.05% + 8 0.11 14 µA
100 k 1 - - 0.05% + 8 0.11 1.5 µA
1000 k 10 - - 0.06% + 8 0.12 150 µA
10 M 100 - - 0.25% + 6 1.5 150 µA
100 M 100 k - - 2% + 2 2.75 320 µA
* Because of the method used to measure resistance, the 100 M (slow) and 300 M(medium and fast)
ranges cannot measure below 3.2 M and 20 M, respectively. "UL" (underload) is shown on the display
for resistances below these nominal points, and the computer interface outputs "+1E-9".