Service Manual
4-6
Table 4-2. Performance Tests for Volts, Diode Test, Ohms, and Frequency Functions (cont)
Function Range (Rate)*** Input
Level
Frequency Display
Min Max
O 3k * short 0.0000 0.0002
1.9 k 1.8988 1.9012
30 k 19 k 18.988 19.012
300 k 190 k 189.88 190.12
3 M 1.9 M 1.8987 1.9013
30 M** 19 M 18.949 19.051
300 M** 190 M 186.2 193.8
Using decades of 1.0:
300 * short 0.00 0.04
100 99.93 100.09
3 k short 0.0000 0.0002
1 k 0.9993 1.0007
30 k 10 k 9.993 10.007
300 k 100 k 99.93 100.07
3 M 1 M 0.9992 1.0008
30 M** 10 M 9.972 10.028
300 M** 100 M 98.0 102.0
F 10/100 kHz 0.1 to 10 V 10 kHz 9.994 10.006
* Use either 2-wire compensation on the 5700A or the relative (REL) mode on the 45.
** Optional test points that can be used if standards are available.
*** All measurement rates are in medium speed unless otherwise specified.