45
Service Manual
4-12
chapter.) In the following procedure, the Fluke 5700A Multifunction Calibrator is used
as the ohms source, with input values of 190 , 1.9 k, 19 k, 190 k, and 1.9 M.
1. If the meter is already in calibration mode, press the
U or T button to select
calibration. Otherwise, select calibration mode first (press Cal Enable button for
three seconds), then select calibration by pressing the
U button six times.
2. Connect the ohms source to the meter in a four-wire configuration. Refer to Figure
4-1.
3. On the 5700A, select 2-wire compensation ON, with external sense. Then program
the 5700A for the output in step 25 of Table 4-5.
4. Edit the meter prompt display to read the same value as the 5700A output display.
5. On the meter, press the
E button to calibrate this step. Then continue with steps
26 through 29 in Table 4-5.
CAL
ENABLE
POWER
AUTO
2ND
MN MX
RATE
dB
REL
REF# LOCAL
THRESH ADDR BAUD
HOLD
LO
HI
COMP
A
A
V
FREQ
V
REF
REMOTE
EXT TRG
SMF MAX
MIN
dB
HOLD
REL
AUTO
mA
mVDCAC
Mk
Hz
UNCAL
mA
mV
DC AC
Hz
Mk
5700A
UUT
COMP
COMP
EX SNS
: ON
EX GRD
: OFF
2-WIRE
COMP
ON
CAUTION
USE CONNECTIONS WITH
EXPOSED PLUG TIPS FOR
THE OHMS FUNCTION ONLY.
HI
HI
LO
LO
HI
SENSE
V
AUX
CURRENT
GUARD GROUND
WIDEBAND
5700A
NC
NC
OUTPUT
V A
45
DUAL DISPLAY MULTIMETER
1000V CAT
FUSE F1
500 mA
F 250V
FUSED
600V CAT
10A
100
mA
V
COM
qb18f.eps
Figure 4-1. Four-Wire Configuration