Getting Started
Battery Status Display
2
2-45
Battery Status Display
To see the charge level of the main units NiMH battery, turn the rotary switch to
SPECIAL FUNCTIONS; then select Battery Status. To see the charge
level of the remotes battery, connect the remote to the main unit; then use ! to
toggle the display.
Conditioning the Battery
The battery status is more accurate when the battery is properly conditioned. The
test tool includes a battery conditioning feature. To condition the batteries, proceed
as follows:
1. Connect the main and remote units together with a calibration module, or with
LIA adapters and an RJ45 patch cord or a cable.
2. Connect ac adapters to both units.
3. Turn on the remote.
4. Press and hold D while turning on the main unit. Release the key when the
display goes blank.
The conditioning function discharges then charges the battery. This may take
several hours, depending on the battery status when conditioning starts. You can
interrupt the conditioning if necessary. The ac power LEDs on the test tool and
remote turn green when conditioning is complete.