Compression Nut
Compression Fitting
Ball Valve
Flow Direction Arrow
Nipple
Step 2 - Remove the Probe from the Retractor Body
Fig. 7.2: Identication of Retractor Parts for the FT3
5. Continue supporting the meter while performing the following steps.
6. If the meter is to be completely removed from the system, disconnect the wiring to the meter.
7. Using a 7/8" wrench, slowly loosen the compression tting approximately one turn or less until
the pressure in the retractor is completely relieved.
8. After relieving the pressure, unscrew the compression tting from the end of the nipple.
Maintenance | 114
Model FT3
Maintenance