Enter the maximum pulse rate (frequency) and press OK (F4).
CAUTION! Maximum pulse rate (frequency) cannot exceed 100 Hz.
The next screen will show:
F1 F2 F3 F4
NOTE! If the ow rate exceeds the maximum pulse rate (frequency), the output will stay at
100 Hz and the FT3 will issue an alarm code.
Alarm Output
To program the Alarm output, press I/O (F1) key from the SET PARAMETERS menu. Then select I/O
(F2) and the screen will show:
F1 F2 F3 F4
Then press OUT (F2) and the screen may show:
F1 F2 F3 F4
Then press NXT (F1) to select the correct alarm and press OK (F4).
Selections are:
Not used
Frequency
HiFloAlm = High Flow Alarm
LoFloAlm = Low Flow Alarm
HiTempAlm = High Temperature Alarm
LoTempAlm = Low Temperature Alarm
System Alm = System Alarm
MaxFreq=100 Hz
CHG OK
MaxFlo=5000 SCFM
CHG OK
SET I/O
INP OUT EXIT
OUT = HiFloAlm
NXT OK