When the output is set to Alarm and there is no alarm condition, the output will be on (0 volts). When
an alarm is active, the output is turned o (24 volts).
F1 F2 F3 F4
Enter the value for the limit by pressing CHG (F1) and then OK (F4). A value of 0 disables the alarm.
NOTE! There is only one output to operate as a pulse output or an alarm output. Both
cannot operate at the same time.
Contact Input Settings
From the SET PARAMETERS menu, press I/O (F1) and then I/O (F2) and then INP (F1) key to select
input. The following menu will display:
F1 F2 F3 F4
Press NXT (F1) until the correct selection is shown and then press OK (F4) to accept the setting.
Selections are:
Not used
Tot Reset = Reset the totalizer
Switch Crv = Switch between calibration curve 1 and curve 2 (only if 2 gas curve ordered)
Press EXIT (F4) repeatedly until you exit programming mode.
FT3 with 2 Gas Curves
This section describes added features to the standard FT3 ow meter when using the 2 gas curves
rmware option. The 2 Gas Curves rmware allows the use of two dierent calibration tables when
running with dierent gases. One of two methods can be used to switch between the two calibration
curves:
1) Use of Contact Input:
When the contact input is programmed for curve switching, an open contact will select curve #1 and a
contact closure will select curve #2.
2) Use of the Keypad:
If the contact input is not programmed for curve switching, pressing F2 and F3 simultaneously will
prompt an operator to manually switch curve upon entering a password and conrming the action by
pressing the appropriate key.
HiFloAlm=500 SCFM
CHG OK
INP=Not Used
NXT OK
Operation | 59