88
464
Incorrect display version (for pre-series only)
Cause: Display software and IG BRAIN are not compatible
Remedy: Update IG BRAIN or display software
Cause: IG BRAIN hardware and display not compatible
Remedy: Replace display or IG BRAIN
465 *
Unknown display version
Cause: Communication between IG BRAIN and display faulty
Remedy: Replace display or IG BRAIN
Remedy: Update software
466
Display could not be recognized
Cause: Communication between IG BRAIN and display faulty
Remedy: Check the ribbon cable between IG BRAIN and display
Cause: Ribbon cable of the display defective
Remedy: Replace display
Cause: IG BRAIN or display defective
Remedy: Replace IG BRAIN or display
467 *
Error during start-up
Cause: Communication error between display and IG BRAIN
Remedy: Remedied automatically via DC reset
Cause: Ribbon cable of the display defective
Remedy: Replace display
Cause: IG BRAIN defective
Remedy: Replace IG BRAIN
468
Memory block for menu entry faulty
Cause: Communication error between display and IG BRAIN
Remedy: Remedied automatically
Cause: Memory error in display
Remedy: Replace display
Cause: Memory error in IG BRAIN
Remedy: Replace IG BRAIN