If you press ‘ENTER’ here, the factory-stage calibration will be restored (calibration, language), the memory
will be erased as well as the programming and you will return to the main configuration screen.
10.4.3.2 Viscosity calibration
When selecting the viscosity option (moving through the menu with the ‘’ and ‘’ key) and pressing
‘ENTER’ the user will access to the following screens, depending on the model of the viscometer:
Model L
Models R and H
The possible spindles’ list to use depends on the model of your viscometer (L, R or H). Thus, in tables 1 to 5
(page 50 and 51) you can see the different available spindles for each model.
Once this field is selected and situated in the list of corresponding spindles, you can select the spindle that
you wish to calibrate using the ‘’ and ‘’ keys.
Once the spindle is selected, go to the “Viscosity” field using the ‘TAB’ key. Pressing ‘ENTER’, accept the field
and introduce the value of the standard oil corresponding to the viscosity calibration. To introduce the data,
use the ‘’ and ‘’ keys to increase or decrease the value of each digit. Then, press ‘TAB’ again to go from
one-digit place to another.
Once the value of the oil is determined, press ‘ENTER’ to continue with the calibration process.
Afterwards, press the ‘ON’ key and the following screen will be shown: