16. ADVANCE L Special aerial spindle selection table
Maximum guideline values in cP (mPa·s)
RPM / SP
L2C
L3C
0.3
100K
400K
0.5
60K
250K
0.6
50K
200K
1
30K
125K
1.5
20K
80K
2
15K
60K
2.5
12K
50K
3
10K
40K
4
7.5K
30K
5
6K
25K
6
5K
20K
10
3K
12K
12
2.5K
10K
20
1.5K
6K
30
1K
4K
50
600
2.5K
60
500
2K
100
300
1.2K
Table 10.
ATTENTION:
K Indicates thousands. Example: 7.8K = 7800
M Indicates millions Example: 1.56M = 1560000
NOTE: It is not recommended to work with viscosity values of less than 15% of the lower part of the selected
scale.