27. ADVANCE H Special spindle selection table
Maximum guideline values in Poise
RPM / SP
TR8
TR9
TR10
TR11
TR12
TR13
0.3
13.6K
66.6K
133.3K
266.6K
133.3K
333.3K
0.5
8K
40K
80K
160k
80K
200K
0.6
6.6K
33.3K
66.6K
133.3K
66.6K
166.6K
1
4K
20K
40K
80K
40K
100K
1.5
2.6K
13.3K
26.6K
53.3K
26.6K
66.6K
2
2K
10K
20K
40K
20K
50K
2.5
1.6K
8K
16K
32K
16K
40K
3
1.3K
6.6K
13.3K
26.6K
13.3K
33.3K
4
1K
5K
10K
20K
10K
25K
5
800
4K
8K
16K
8K
20K
6
666
3.30K
6.6K
13.3K
6.6K
16.6K
10
400
2K
4K
8K
4K
10K
12
333
1.6
3.3K
6.6K
3.3K
8.3K
20
200
1K
2K
4K
2K
5K
30
133
666
1.3K
2.6K
1.3K
3.3K
50
80
400
800
1.6K
800
2K
60
66
333
666
1.3K
666
1.6K
100
40
200
400
800
400
1K
Table 21.
ATTENTION:
K Indicates thousands. Example: 7.8K = 7800
M Indicates millions Example: 1.56M = 1560000
NOTE: It is not recommended to work with viscosity values of less than 15% of the lower part of the selected
scale.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fungilab VISCOLED ADVANCE Series and is the answer not in the manual?

Fungilab VISCOLED ADVANCE Series Specifications

General IconGeneral
BrandFungilab
ModelVISCOLED ADVANCE Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals