Noise emitted: Set less than 70 dB (A)
Vibrations emitted: Less than 2.5 m/s²
Certifications:
Conforms to CE Standards:
EN 61010-1:2010 Safety requirements for electrical equipment, for measurement, control
and laboratory use.
EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use
Directive RoHS: 2011/65/UE + 2014/IUE a 2014/6/UE +2014/8/UE a 2014/16/UE
Notice to customers:
The product is made up of various components and various materials that
must be recycled or, failing that, deposited in the corresponding debris
removal sites when the product's life has been completed or when
otherwise it is necessary to dispose of it. To do this, the end user who
acquires the product must know the current regulations of each
municipality and / or locality based on the waste electrical and electronic
equipment. The user who acquires this product must be aware of and
responsible for the potential effects of the components on the environment
and human health as a result of the presence of hazardous substances.
Never place the product in a conventional container of citizen scope if a
previous dismantling and knowledge of the components that incorporates.
If you do not know the procedure to follow, consult with the city council for
more information.
6. Conditions for use
-
Indoor use
-
Maximum altitude 2000 m.
-
Surrounding temperature range: from +5 to 40ºC.
-
The equipment temperature must be kept above the dew point so moisture doesn't condense on
or in it. The power source fluctuations should not surpass 10% of the nominal voltage
-
Installation category II
-
Pollution level II
7. Maintenance
Always clean all of the parts after each use! Clean and dry the spindles and the spindle guard well.
Make sure that there is not any sample remaining, especially in the delicate zones such as the
spindle connector.
Use detergents or solvents to clean the spindles and the protector:
-
For cleaning food samples, use lukewarm water and if necessary, use soft household
detergents
-
Other solvents that generally provide good results are acetone, gasoline, or any solvent
with a high percentage of alcohol
-
For the use of any other solvent, make sure that it does not corrode the spindles or the
protector. The spindles are made in AISI 316.
Warning: Handle the volatile and flammable solvents with proper cautions. It is the
user’s responsibility to establish safety conditions at work.
Check regularly the spindle’s thread and the viscometer shaft.
During the viscometer’s lifespan, regular maintenance is important. FUNGILAB recommends annual
check-ups by the technical service of your local distributor.