18. ADVANCE L LCP Adaptor table
Maximum guideline values in cP (mPa·s)
RPM/SP
LCP
0.3
2K
0.5
1.2K
0.6
1K
1
600
1.5
400
2
300
2.5
240
3
200
4
150
5
120
6
100
10
60
12
50
20
30
30
20
50
12
60
10
100
6.0
Table 12.
ATTENTION:
K Indicates thousands. Example: 7.8K = 7800
M Indicates millions Example: 1.56M = 1560000
NOTE: It is not recommended to work with viscosity values of less than 15% of the lower part of the selected
scale.