21. ADVANCE R standard spindle selection table
Maximum guideline values in cP (mPa·s)
RPM / SP
R1
R2
R3
R4
R5
R6
R7
0.3
33.3K
133.3K
333.3K
666.6K
1.3M
3.33M
13.3M
0.5
20K
80K
200K
400K
800K
2M
8M
0.6
16.6K
66.6K
166.6K
333.3K
666.6K
1.6M
6.6M
1
10K
40K
100K
200K
400K
1M
4M
1.5
6.6K
26.6K
66.6K
133.3K
266.6K
666.6K
2.6M
2
5K
20K
50K
100K
200K
500K
2M
2.5
4K
16K
40K
80K
160K
400K
1.6M
3
3.3K
13.3K
33.3K
66.6K
133.3K
333.3K
1.3M
4
2.5K
10K
25K
50K
100K
250K
1M
5
2K
8K
20K
40K
80K
200K
800K
6
1.6K
6.6K
16.6K
33.3K
66.6K
166.6K
666.6K
10
1K
4K
10K
20K
40K
100K
400K
12
833
3.3K
8.3K
16.6K
33.3K
83.3K
333.3K
20
500
2K
5K
10K
20K
50K
200K
30
333
1.3K
3.3K
6.6K
13.3K
33.3K
133.3K
50
200
800
2K
4K
8K
20K
80K
60
166
660
1.6K
3.3K
6.6K
16.6K
66.6K
100
100
400
1K
2K
4K
10K
40K
Table 15.
ATTENTION:
K Indicates thousands. Example: 7.8K = 7800
M Indicates millions Example: 1.56M = 1560000
NOTE: It is not recommended to work with viscosity values of less than 15% of the lower part of the selected
scale.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fungilab VISCOLED ADVANCE Series and is the answer not in the manual?

Fungilab VISCOLED ADVANCE Series Specifications

General IconGeneral
BrandFungilab
ModelVISCOLED ADVANCE Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals