19. ADVANCE L special Vane spindle selection table
Maximum guideline values in cP (mPa·s)
RPM / SP
V71
V72
V73
V74
V75
0.3
8.1K
34.6K
167K
1.6M
721K
0.5
4.9K
20.8K
100K
1M
433K
0.6
4K
17.3K
83.5K
848K
360K
1
2.4K
10.4K
50.1K
508K
216K
1.5
1.6K
6.93K
33.4K
339K
144K
2
1.2K
5.2K
25K
254K
108K
2.5
982.2
4.1K
20K
203K
86.6K
3
818.5
3.4K
16.7K
169K
72.1K
4
613.9
2.6K
12.5K
127K
54.1K
5
491.1
2k
10K
101K
43.3K
6
409.2
1.7K
8.3K
84.8K
36K
10
245.5
1K
5K
50.8K
21.6K
12
204.6
867
4.1K
42.4K
18K
20
122.7
520.2
2.5K
25.4K
10.8K
30
81.8
346.8
1.6K
16.9K
7.2K
50
49.1
208
1K
10.1K
4.3K
60
40.9
173.4
835.7
8.4K
3.6K
100
24.5
100
501.4
5K
2.1K
Table 13.
ATTENTION:
K Indicates thousands. Example: 7.8K = 7800
M Indicates millions Example: 1.56M = 1560000
NOTE: It is not recommended to work with viscosity values of less than 15% of the lower part of the selected
scale.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fungilab VISCOLED ADVANCE Series and is the answer not in the manual?

Fungilab VISCOLED ADVANCE Series Specifications

General IconGeneral
BrandFungilab
ModelVISCOLED ADVANCE Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals